Simple method for estimating linear attenuation coefficients from x-ray diffraction tomography data

Citation
Km. Pavlov et al., Simple method for estimating linear attenuation coefficients from x-ray diffraction tomography data, REV SCI INS, 72(3), 2001, pp. 1918-1920
Citations number
11
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
REVIEW OF SCIENTIFIC INSTRUMENTS
ISSN journal
00346748 → ACNP
Volume
72
Issue
3
Year of publication
2001
Pages
1918 - 1920
Database
ISI
SICI code
0034-6748(200103)72:3<1918:SMFELA>2.0.ZU;2-Y
Abstract
In this article, we demonstrate how the spatial distribution of x-ray linea r attenuation coefficients within an object can be estimated from x-ray dif fraction tomography data. The experimental arrangement to achieve this expl oits a position sensitive detector and an analyzer crystal. The quality of reconstructed maps of linear attenuation coefficients is comparable to resu lts obtained from conventional transmission computed tomography. (C) 2001 A merican Institute of Physics.