An optimized fast protocol for magnetic resonance imaging of the temporomandibular joint

Citation
E. Honda et al., An optimized fast protocol for magnetic resonance imaging of the temporomandibular joint, DENTOMAX R, 30(2), 2001, pp. 126-130
Citations number
14
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging
Journal title
DENTOMAXILLOFACIAL RADIOLOGY
ISSN journal
0250832X → ACNP
Volume
30
Issue
2
Year of publication
2001
Pages
126 - 130
Database
ISI
SICI code
0250-832X(200103)30:2<126:AOFPFM>2.0.ZU;2-U
Abstract
Objectives: To provide a fast protocol for MR imaging of the TMJ with high contrast resolution of both soft tissue and joint fluid. Methods: A fast turbo-spin echo (TSE) pulse sequence was developed. The new technique was compared with TIW conventional spin echo (CSE) and T2W TSE s equences in imaging 57 TMJs in 50 patients. Disc location and contour and b one delineation were assessed on three images of each TMJ by one observer. Presence and location of intra-articular fluid were evaluated on both stand ard T2W and new TSE images. Image quality was scored on a 3 point-scale by one observer. Results: The net image acquisition time for one plane was 2 min with the ne w TSE sequence compared with approximately 5 - 10 min for TIW and 4 min for T2W. The new sequence provided equivalent diagnostic information to a comb ination of TIW CSE for disc position and contour and bone contour and T2W T SE sequence for joint fluid. Conclusions: The examination time for MRI of the TMJ can be considerably re duced with the new fast TSE pulse sequence without compromising image quali ty.