Spatial resolution in Micromegas detectors

Citation
J. Derre et al., Spatial resolution in Micromegas detectors, NUCL INST A, 459(3), 2001, pp. 523-531
Citations number
2
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
459
Issue
3
Year of publication
2001
Pages
523 - 531
Database
ISI
SICI code
0168-9002(20010301)459:3<523:SRIMD>2.0.ZU;2-8
Abstract
The performance of a telescope of Micromegas detectors has been studied in a pion beam at the CERN PS. With a gas filling of CF4 and 20% isobutane and with a strip pitch of 100 mum an accuracy of 14 +/- 3 mum on the spatial r esolution has been measured at normal incidence. A simulation demonstrates that the resolution is limited by the size of the holes of the mesh of the detector and could be reduced to 11 mum in the same conditions with smaller holes. Even further improvement down to 8.5 mum is feasible for the same g as with an optimized 75 mum strip pitch. (C) 2001 Elsevier Science B.V. All rights reserved.