Y. Nakane et Y. Sakamoto, Measurement of absorbed dose distributions in a plastic phantom irradiatedby 40-and 65-MeV quasi-monoenergetic neutrons, NUCL INST A, 459(3), 2001, pp. 552-564
In order to validate dose evaluation methods for radiation shielding design
of accelerator facilities, absorbed dose distributions in a plastic phanto
m of 30 x 30 x 30 cm(3) slab were measured for 40- and 65-MeV quasi-monoene
rgetic neutrons with a tissue-equivalent proportional counter and a tissue-
equivalent ionization chamber. The measured distributions were compared wit
h the calculated ones from the energy deposition of charged particles and n
eutron energy spectra by using the Monte Carlo codes, HETC-3STEP and MORSE-
CG/KFA, respectively. As a result, it was revealed that the calculation met
hod used in the present work was applicable to absorbed dose evaluation for
intermediate energy neutrons with high accuracy. (C) 2001 Elsevier Science
B.V. All rights reserved.