Double-undulator tests of a diamond monochromator at the Advanced Photon Source

Citation
Pb. Fernandez et al., Double-undulator tests of a diamond monochromator at the Advanced Photon Source, NUCL INST A, 459(1-2), 2001, pp. 347-353
Citations number
11
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
459
Issue
1-2
Year of publication
2001
Pages
347 - 353
Database
ISI
SICI code
0168-9002(20010221)459:1-2<347:DTOADM>2.0.ZU;2-M
Abstract
We have tested the thermal performance of a water-cooled diamond double-cry stal monochromator under twice the typical undulator thermal load at the Ad vanced Photon Source (APS). With two 3.3-cm-period undulators installed in series in the APS sector 1 straight section, the maximum power and normal p ower density incident on the first crystal were 705 W and 312 W/mm(2), resp ectively. The maximum power and surface power density absorbed by the coole d diamond were 140 W and 16.7 W/mm(2), respectively. The monochromator perf ormed well under this large power load, demonstrating the viability of usin g water-cooled diamonds for high-intensity synchrotron radiation beamlines. (C) 2001 Elsevier Science B.V. All rights reserved.