Humidity sensor using cross-linked copolymers containing viologen moiety

Citation
Ms. Gong et al., Humidity sensor using cross-linked copolymers containing viologen moiety, SENS ACTU-B, 73(2-3), 2001, pp. 185-191
Citations number
12
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
SENSORS AND ACTUATORS B-CHEMICAL
ISSN journal
09254005 → ACNP
Volume
73
Issue
2-3
Year of publication
2001
Pages
185 - 191
Database
ISI
SICI code
0925-4005(20010310)73:2-3<185:HSUCCC>2.0.ZU;2-E
Abstract
1-[3-(2-Methacryloxyerhyl dimethyl ammono)propyl] 1 '-propyl-4,4'-bipyridin ium tribromide (MDAPBT) was selected as a humidity sensing monomer. n-Butyl acrylate (n-BA) and 2-(N,N'-dimethylaminoethyl)methacrylate (DAEMA) were u sed as a comonomer. The major ingredient of the humid membrane is the copol ymers of MDAPBT with n-BA and DAEMA. The cross-linked copolymer of MDAPBT/n -BA/DAEMA = 40/30/30 showed average resistance of 1000, 52 and 3.1 K Omega at 30, 60 and 90% relative humidity (RH), respectively. Temperature depende nce, frequency dependence and response time were also measured. The tempera ture dependence at 15-40 degreesC is -0.42 to -0.45% RH degreesC(-1) and th e hysteresis falls on the range of +/-2% RH. The humidity-sensitive charact eristics of the sensor did not change even after 20 min when the sensor was soaked in water. (C) 2001 Elsevier Science B.V. All rights reserved.