S. Raut et al., ULTRA-RAPID PREPARATION OF MILLIGRAM QUANTITIES OF THE PURIFIED POLYPEPTIDE-CHAINS OF HUMAN FIBRINOGEN, Journal of chromatography B. Biomedical applications, 660(2), 1994, pp. 390-394
Citations number
19
Categorie Soggetti
Chemistry Analytical
Journal title
Journal of chromatography B. Biomedical applications
In order to study the epitopes in fibrin towards which monoclonal anti
bodies are directed we needed the pure individual polypeptide chains o
f human fibrinogen in reasonable quantity. We report here a simplified
, rapid method of separation of high-purity human fibrinogen chains. F
ollowing reduction and S-carboxymethylation of human fibrinogen, the s
ample was injected directly onto a column of the polymeric reversed-ph
ase perfusion packing POROS 20-R2, and the chains were completely reso
lved in less than 3 min at a flow-rate of 10 ml/min. The capacity was
equivalent to that of a similar sized conventional silica-based column
. However the throughput was approximately five to ten times as high.
The column was durable and robust in day-to-day use.