A pick-up system for precise magnetization measurements in pulsed field fac
ilities is presented. Samples with diameters up to 5 mm can be measured (di
ameter of the pick-up coil is 6.8 mm, outer diameter of the pick-up system
is 11.8 mm). The optimized geometry of the concentric pick-up and compensat
ion coils reduces the influence of a sample displacement on the measured si
gnal. The measured signal changes by less than 0.2% if the position of a sp
herical sample of 4.3 mm diameter is shifted by +/- 2mm along the axis of t
he coils. An additional compensation coil is used, to correct the differenc
e between the winding areas of pick-up and compensation coils. This correct
ion as well as the signal of the field measuring coils depend only weakly o
n an axial displacement of the pick-up system relatively to the center of t
he magnetic field. The calibration of the field measuring coils is performe
d by the metamagnetic transition of a MnF2 single crystal (H = 9.3 T at T =
4.2 K). The pick-up system is used for magnetization measurements in pulse
d fields up to 50T at temperatures between 2 and 300K in the high magnetic
field laboratory 'Hochfeldlabor Dresden'. (C) 2001 Elsevier Science B.V. Al
l rights reserved.