Design and testing of high sensitivity microreceiver coil apparatus for nuclear magnetic resonance and imaging

Citation
Da. Seeber et al., Design and testing of high sensitivity microreceiver coil apparatus for nuclear magnetic resonance and imaging, REV SCI INS, 72(4), 2001, pp. 2171-2179
Citations number
46
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
REVIEW OF SCIENTIFIC INSTRUMENTS
ISSN journal
00346748 → ACNP
Volume
72
Issue
4
Year of publication
2001
Pages
2171 - 2179
Database
ISI
SICI code
0034-6748(200104)72:4<2171:DATOHS>2.0.ZU;2-I
Abstract
We report the design and testing of a nuclear magnetic resonance (NMR) micr ocoil receiver apparatus, employing solenoidal microreceiver coils of dimen sions of tens to hundreds of microns, using applied field of 9 T (proton re sonance frequency 383 MHz). For the smallest receiver coils we attain sensi tivity sufficient to observe proton NMR with signal to noise (S/N) one in a single scan applied to a similar to 10 mum(3) (10 fl) water sample, contai ning 7x10(11) total proton spins. We also test the dependence of the S/N on important coil parameters, including coil composition and resistivity, tur n spacing, and lead lengths. (C) 2001 American Institute of Physics.