M. Vahlensieck et al., Magnetization transfer contrast (MTC): Optimization of the off-resonance and on-resonance MTC methods at 0.5 and 1.5 T, BIOMED TECH, 46(1-2), 2001, pp. 10-17
Aim: To compare technical parameters and clinically relevant aspects of on-
and off-resonance MTC sequences in mid- and high-field MR systems.
Material and Methods: Both on- and off-resonance techniques were combined w
ith an FFE sequence using 0.5 and 1.5 Tesla superconducting systems. Parame
ters were systematically measured by scanning a cadaveric knee joint. Signa
l-to-noise ratios and MT ratios for fat, cartilage and reference solution (
copper sulphate) were determined. Minimal TR and the energy absorption rate
were also compared.
Results: The MT effect at 1.5 T was more pronounced. However, using optimiz
ed parameters, clinically adequate MT contrast was achieved with both techn
iques and both MT units. The most important parameters for the off-resonanc
e method are pulse angle and off-centre frequency; for the on-resonance met
hod, pulse angle and number of composite pulse elements. Energy absorption
was below 2 W/kg. Minimal TR was prolonged by up to 400 %.
Conclusion: In order to produce MTC images, optimized parameters should be
applied. Using optimized pulse parameters, adequate MTC imaging is achievab
le with mid- and high-field systems using on- and off-resonance techniques.
To ensure comparability of MTC studies, the pulse parameters need to be gi
ven, and, ideally, standardized.