Reference doses in dental radiodiagnostic facilities

Citation
L. Gonzalez et al., Reference doses in dental radiodiagnostic facilities, BR J RADIOL, 74(878), 2001, pp. 153-156
Citations number
20
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Journal title
BRITISH JOURNAL OF RADIOLOGY
ISSN journal
00071285 → ACNP
Volume
74
Issue
878
Year of publication
2001
Pages
153 - 156
Database
ISI
SICI code
Abstract
The present work describes an experiment undertaken using thermoluminescent dosemeters to obtain a local reference dose level in orthopantomography, b ased on patient measurements in 11 dentistry installations. In addition, a critical audit of the dose recommended by the International Atomic Energy A gency for intraoral X-ray diagnostics is performed, based on data gathered from over 300 intraoral X-ray facilities. The provisional local reference l evels proposed are 0.7 mGy entrance surface dose at the occipital region fo r orthopantomography and 3.5 mGy entrance surface dose for intraoral radiol ogy.