Re. Chrien et al., CALIBRATION AND OPERATION OF A NEUTRON TIME-OF-FLIGHT SCINTILLATOR ARRAY ON NOVA, Fusion engineering and design, 34-5, 1997, pp. 577-580
The Tion diagnostic is a 1020-channel scintillator array on the Nova l
aser at Livermore. It measures the arrival time of single D-D or D-T n
eutrons at 27 m from the target with time resolution of 4 ns. The effe
ct of neutron scattering has been obtained from Monte-Carlo neutron tr
ansport simulations and the scattered response is convolved with a Gau
ssian source to fit the data. The data analysis has been checked using
targets which produce roughly equal numbers of D-T and D-D counts. Th
e temperature obtained from Tion and from a current-mode neutron time-
of-flight detector are in agreement for neutron yields above 1 x 10(9)
. (C) 1997 Elsevier Science S.A.