In vitro radiobiological studies with ions require an extracted beam in air
, uniformly spread over an area of at least 1 cm(2) characterised by a good
monochromaticity. Moreover an accurate fluence monitoring is needed togeth
er to an accurate off line calibration, taking into account the measurement
geometry, to determine the absorbed dose. Finally energy, and then LET, fo
llowing every traversed material, should be determined. In this framework w
e set up a facility at the 3 MV Tandem HVC TTT-3 accelerator of the Univers
ity of Naples "Federico II" using light ions up to Z = 6 and energy below 0
.6 MeV n(-1). Survival measurements of the V79 cell line following irradiat
ion with beryllium, boron and carbon showed a good agreement with data from
literature. (C) 2001 Elsevier Science B.V. All rights reserved.