In this study a complete set of dosimetric data is presented for the high d
ose rate (HDR) source from Amersham used in the Buchler remote afterloading
HDR unit. These data have been calculated by means of the Monte Carlo simu
lation code GEANT taking into account the detailed geometry of the source.
Absolute dose rate distributions in water were calculated around this sourc
e and are presented as conventional 2D Cartesian look-up tables. All dosime
tric quantities recommended by the AAPM Task Group 43 report have been calc
ulated. Quantities determined are: dose rate constant, radial dose function
, anisotropy Function, anisotropy factor and anisotropy constant. The dose
rate distributions of the Buchler HDR source are compared with those of oth
er HDR sources used in brachytherapy, showing that the differences are larg
e in zones near the long source axis due to oblique filtration. These Monte
Carlo simulated data in water can be used for clinical applications.