Solventless sample preparation procedure for organophosphorus pesticides analysis using solid phase microextraction and on-line supercritical fluid extraction/high performance liquid chromatography technique

Citation
Sh. Salleh et al., Solventless sample preparation procedure for organophosphorus pesticides analysis using solid phase microextraction and on-line supercritical fluid extraction/high performance liquid chromatography technique, ANALYT CHIM, 433(2), 2001, pp. 207-215
Citations number
25
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICA CHIMICA ACTA
ISSN journal
00032670 → ACNP
Volume
433
Issue
2
Year of publication
2001
Pages
207 - 215
Database
ISI
SICI code
0003-2670(20010411)433:2<207:SSPPFO>2.0.ZU;2-T
Abstract
Organophosphorus pesticides spiked in water sample solutions were extracted with solid phase microextraction (SPME) technique and were desorbed by sup ercritical fluid carbon dioxide (SFCO2) before on-line introduction into hi gh performance liquid chromatography system (HPLC). All of the 5-mul SFCO2 injected into the HPLC system were dissolved into the methanol/water (80/20 ) mobile phase without affecting the baseline of the UV detector signal. Pr eliminary experiments were performed in order to optimize the extraction co nditions of pesticides from spiked water solutions with the SPME technique. Subsequently, desorption conditions of the pesticides from the SPME fiber coating using SFCO2 as the desorption medium were optimized. (C) 2001 Elsev ier Science B.V. All rights reserved.