Nc. Dobrolet et Ja. Ettedgui, Changes in coil morphology following transcatheter coil occlusion of the patent ductus arteriosus using a modified snare technique, CATHET C IN, 52(4), 2001, pp. 504-509
We observed a decrease in length of the Gianturco coils following transcath
eter occlusion of the patent ductus arteriosus (PDA), Coil length was measu
red on chest radiograph within 24 hr of coil placement and compared to the
length at the time of follow-up, Echocardiograms were also reviewed for evi
dence of duct recanalization. Twenty-seven patients met inclusion criteria,
The median time to follow-up was 5 months (1-12 months), Twenty-four out o
f 27 (89%) patients had a decrease in coil length by 1.9 +/- 1.1 mm (P < 0.
01). This was an average decrease of 16%. One patient had an increase in le
ngth and two patients had no change in coil length. Nineteen out of 27 pati
ents had echocardiograms, Despite the change in coil length, there was no e
vidence of flow acceleration in the pulmonary artery or descending aorta. N
o patients had evidence of duct recanalization, <(c)> 2001 Wiley-Liss, Inc.