DETERMINATION OF CHLOROPHENOLS IN DRINKING-WATER SAMPLES AT THE SUBNANOGRAM PER MILLILITER LEVEL BY GAS-CHROMATOGRAPHY WITH ATOMIC-EMISSIONDETECTION

Citation
Mi. Turnes et al., DETERMINATION OF CHLOROPHENOLS IN DRINKING-WATER SAMPLES AT THE SUBNANOGRAM PER MILLILITER LEVEL BY GAS-CHROMATOGRAPHY WITH ATOMIC-EMISSIONDETECTION, Journal of chromatography, 683(1), 1994, pp. 21-29
Citations number
29
Categorie Soggetti
Chemistry Analytical
Journal title
Volume
683
Issue
1
Year of publication
1994
Pages
21 - 29
Database
ISI
SICI code
Abstract
The use of an atomic emission detector following a process of preconce ntration of drinking water samples by a factor of 1500:1 allows the hi ghly selective determination of chlorophenols present in samples below the maximum limit of 0.5 ng/ml set by international regulations. The preconcentration of the samples is carried out using 0.25-g commercial graphitized carbon cartridges without the need for sample derivatizat ion prior to solid-phase extraction.