A. Moreda-pineiro et al., Chemometrics approaches for the study of systematic error in inductively coupled plasma atomic emission spectrometry and mass spectrometry, J ANAL ATOM, 16(4), 2001, pp. 350-359
Systematic errors observed when using inductively coupled plasma atomic emi
ssion spectrometry (ICP-AES) and mass spectrometry (ICP-MS) for the multi-e
lement determination in acid digests of environmental samples (tea leaves)
were evaluated. Two chemometric approaches, experimental design and princip
al component analysis, were used in order to establish the errors associate
d with each stage of the analytical method: sample digestion (effect of the
"number of acid digestions") and measurement step (effect of the "number o
f replicates" and the "calibration"). The elements under study were Co, Cr,
Cs, Cu, Ni, Pb, Rb and Ti by ICP-MS, and Ba, Ca, Fe, Mg, Mn, Sr and Zn by
ICP-MS and ICP-AES. Flame atomic absorption spectrometry was used for compa
rative purposes. A Chinese tea certified reference material with certified
concentration for most of the elements was employed as the sample matrix. V
ariance estimation was made from ANOVA outputs from a full factorial design
(FFD) 4(1) x 4(1) x 2(1).