J. Kampmeier et al., Prospective and randomized comparison of frequency doubling perimetry vs standard automated perimetry in patients with glaucoma, KLIN MONATS, 218(3), 2001, pp. 157-167
Aim: To compare frequency doubling perimetry (FDP) versus standard automate
d perimetry in glaucoma. To evaluate the reproducibility of both methods.
Patients and Methods: All the patients were tested (full threshold test) tw
ice with both FDP (N-30) and Humphrey Field Analyzer II Model 750 (HFA) (30
-2 procedures) in random sequence, within one day. The parameters mean devi
ation (MD), pattern standard deviation (PSD) and measured thresholds per qu
adrant and center areas were evaluated for FDP/HFA comparison in 45 glaucom
a patients. The same parameters were used for testing the reproducibility o
f FDP (n = 46) and HFA (N = 39). Additionally, the center and each quadrant
area were checked for any deviation in the total deviation probability plo
ts. For statistical analysis Kappa coefficients and Bland/Altman plots were
used.
Results: An average MD (FDP vs HFA) of -7.3+/-5.8 dB vs -8.9 +/- 7.6 dB and
PSD of 6.9 +/- 2.4 dB vs 6.5 +/- 3.6 dB were found. Kappa coefficients den
ote marginal accordance (kappa: 0.11-0.38) for area deviations. In a total
of 225 areas HFA detected 191 deviations and FDP 165. HFA measured more neg
ative deviation compared with FDP in the case of MD < -15 dB. The reproduci
bility was 0.98 (MD) and 0.92 (PSD) for FDP and 0.98 (MD) and 0.95 (PSD) fo
r HFA (95% confidence interval).
Conclusions: There was great conformity between FDP and HFA in glaucoma pat
ients. HFA detected more deviations in the total deviation probability plot
s than FDP. There was a high reproducibility of both methods. FDP is an app
ropriate tool for detecting visual field loss in glaucoma patients.