In this paper we describe an experimental set-up for simultaneous measureme
nts of thermopower and electrical resistivity at temperatures from 100 K to
1300 K. Optimal configuration of electrodes and original mechanical contac
ts of thermocouples, current leads and potential probes with the sample mak
e it possible to measure a large variety of materials and result in greater
flexibility with respect to the sample form and dimensions. Both bulk and
thin film samples with resistances in the range from 0 Omega up to 200 k Om
ega (1 G Omega in case of the resistivity measurement) can be investigated.
Precision and high reliability of the system have been proven during more
than three years of use. The resistivity and thermopower of pure Pb, Cu and
Ni, and a Cr-Si thin film composite are presented as test materials to dem
onstrate the possibilities and accuracy of this experimental set-up.