Experimental set-up for thermopower and resistivity measurements at 100-300 K

Citation
At. Burkov et al., Experimental set-up for thermopower and resistivity measurements at 100-300 K, MEAS SCI T, 12(3), 2001, pp. 264-272
Citations number
15
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
MEASUREMENT SCIENCE & TECHNOLOGY
ISSN journal
09570233 → ACNP
Volume
12
Issue
3
Year of publication
2001
Pages
264 - 272
Database
ISI
SICI code
0957-0233(200103)12:3<264:ESFTAR>2.0.ZU;2-V
Abstract
In this paper we describe an experimental set-up for simultaneous measureme nts of thermopower and electrical resistivity at temperatures from 100 K to 1300 K. Optimal configuration of electrodes and original mechanical contac ts of thermocouples, current leads and potential probes with the sample mak e it possible to measure a large variety of materials and result in greater flexibility with respect to the sample form and dimensions. Both bulk and thin film samples with resistances in the range from 0 Omega up to 200 k Om ega (1 G Omega in case of the resistivity measurement) can be investigated. Precision and high reliability of the system have been proven during more than three years of use. The resistivity and thermopower of pure Pb, Cu and Ni, and a Cr-Si thin film composite are presented as test materials to dem onstrate the possibilities and accuracy of this experimental set-up.