Hr. Simonsen et al., International comparison of He-Ne lasers stabilized with I-127(2) at lambda approximate to 543 nm (December 1999) - Northern European lasers, METROLOGIA, 37(6), 2000, pp. 709-714
This second international comparison between He-Ne lasers at 543 nm exhibit
s similar results to the first international comparison. A standard uncerta
inty of 4 kHz was obtained against 2 kHz in 1993, while the average frequen
cy difference between the lasers was -1.0 kHz against 14.2 IcHz in 1993. Th
e experimental data indicate however that a standard uncertainty below 1 Ic
Hz can be expected between two lasers. The frequency difference between the
two lasers clearly lies within the limits (2.5 parts in 10(11)) given in t
he revised practical realization of the definition of the metre (1997) for
the 633 nm He-Ne laser.