We describe a patient whose presentation of ICD generator failure was excru
ciating shoulder pain lasting 3 hours. This painful episode prompted evalua
tion and subsequent generator replacement. Destructive analysis of the expl
anted device revealed a short circuit of three battery filter capacitors, w
hich resulted in the battery "dumping" its full energy very rapidly, since
the impedance across the battery terminal was < 1 ohm. The duration of the
painful episode was equal to the estimated ICD battery life under this low
impedance condition.