Preconcentration and separation with Amberlite XAD-4 resin; determination of Cu, Fe, Pb, Ni, Cd and Bi at trace levels in waste water samples by flame atomic absorption spectrometry

Citation
A. Uzun et al., Preconcentration and separation with Amberlite XAD-4 resin; determination of Cu, Fe, Pb, Ni, Cd and Bi at trace levels in waste water samples by flame atomic absorption spectrometry, TALANTA, 54(1), 2001, pp. 197-202
Citations number
27
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
TALANTA
ISSN journal
00399140 → ACNP
Volume
54
Issue
1
Year of publication
2001
Pages
197 - 202
Database
ISI
SICI code
0039-9140(20010330)54:1<197:PASWAX>2.0.ZU;2-5
Abstract
A method for the preconcentration of Cu, Fe, Pb, Ni, Cd and Bi as their die thyldithiocarbamate chelates was proposed using a column filled with Amberl ite XAD-4 resin. The retained analytes on the resin were recovered with a s mall volume of acetone. The metal ions in the effluent were determined by a flame atomic absorption spectrometer. Different factors including pH of sa mple solution, sample volume, amount of XAD-4 resin, amount of ligand, elue nt volume and matrix effects for preconcentration were examined. The recove ries fur the analytes under the optimum working conditions were higher than 95%. The relative standard deviations of the determinations were below 9%. The limits of detection (3 s, n = 20) for analytes were found to be betwee n 4 and 23 mug l(-1). The proposed method was applied to the analysis of so me waste waters from the organized industrial region of Kayseri. (C) 2001 E lsevier Science B.V. All rights reserved.