Ar. Hubbard et al., Standardisation of factor VIII and von Willebrand factor in plasma: Calibration of the 4th International Standard (97/586), THROMB HAEM, 85(4), 2001, pp. 634-638
The 4th International Standard (IS) Factor VIII/von Willebrand Factor (FVII
I/VWF) plasma was calibrated in 25 laboratories by assay against the 3rd IS
plasma and fresh normal plasma pools. Five parameters were measured, FVIII
:coagulant activity (FVIII:C), FVIII:Antigen (FVIII:Ag), VWF:Antigen (VWF:A
g), VWF:Ristocetin Cofactor (VWF:RCof), and a new parameter, VWF:collagen b
inding (VWF:CB). Mean potency estimates for the 4th IS, calculated relative
to the 3rd IS, were significantly greater than the mean estimates calculat
ed relative to the fresh normal pools by 15, 14 and 20% respectively for FV
III:C, VWF:Ag and VWF:RCof, These results indicate a drift in the Internati
onal Unit away from the fresh plasma unit. Partial rectification of this dr
ift was achieved by assigning the mean of the estimates calculated relative
to the 3rd IS and the fresh plasma pools, i.e. FVIII:C 0.57 IU/ampoule, VW
F:Ag 0.79 IU/ampoule and VWF:RCof 0.73 IU/ampoule. This represents a shift
in the IU between the 3rd and 4th IS of 7.5% for FVIII:C, 7% for VWF:Ag and
10% for VWF:RCof. Mean estimates of FVIII:Ag relative to the 3rd IS and th
e fresh normal pools agreed to give an assigned value of 0.89 IU/ampoule, E
xcessive inter-laboratory variability and a low number of estimates (n = 6)
precluded the assignment of a potency for VWF:CB, The 4th IS Factor VIII/V
WF plasma (97/586) was established in October 1998.