Silver(I) ion-selective membrane based on Schiff base-p-tert-butylcalix[4]arene

Citation
Rk. Mahajan et al., Silver(I) ion-selective membrane based on Schiff base-p-tert-butylcalix[4]arene, ANALYST, 126(4), 2001, pp. 505-507
Citations number
12
Categorie Soggetti
Chemistry & Analysis","Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYST
ISSN journal
00032654 → ACNP
Volume
126
Issue
4
Year of publication
2001
Pages
505 - 507
Database
ISI
SICI code
0003-2654(200104)126:4<505:SIMBOS>2.0.ZU;2-3
Abstract
A PVC membrane electrode for silver(I) ion based on Schiff base-p-tert-buty lcalix[4]arene is reported. The electrode works well over a wide range of c oncentration (1.0 x 10(-5)-1.0 x 10(-1) mol dm(-3)) with a Nernstian slope of 59.7 mV per decade. The electrode shows a fast response time of 20 s and operates in the pH range 1.0-5.6. The sensor can be used for more than 6 m onths without any divergence in the potential. The selectivity of the elect rode was studied and it was found that the electrode exhibits good selectiv ity for silver ion over some alkali, alkaline earth and transition metal io ns. The silver ion-selective electrode was used as an indicator electrode f or the potentiometric titration of silver ion in solution using a standard solution of sodium chloride; a sharp potential change occurs at the end-poi nt. The applicability of the sensor to silver(I) ion measurement in water s amples spiked with silver nitrate is illustrated.