A. Ceresa et al., Influence of key parameters on the lower detection limit and response function of solvent polymeric membrane ion-selective electrodes, J ELEC CHEM, 501(1-2), 2001, pp. 70-76
The lower detection limit of conventional polymeric membrane ion-selective
electrodes (ISEs) is given by the release of measuring ions from the sensor
membrane into the aqueous sample. It can be greatly improved by using inne
r solutions of appropriate composition. Recently, steady-state model calcul
ations have been used to predict the influence of various parameters on the
response function and the lower detection limit. Among these, effects of t
he composition of the inner solution, the presence of interfering ions in t
he sample, the use of a flow-through cell, the polymer content and thicknes
s of the ISE membrane, and the concentration of the membrane components hav
e been investigated in the present work. All observed trends agree frilly w
ith the theoretically predicted ones. (C) 2001 Elsevier Science B.V. Ali ri
ghts reserved.