Background and Purpose: Electrodes used by the Dornier MFL 500 lithotripter
have a specified lifespan, after which, the manufacturer recommends replac
ing the electrode before continuing the treatment, The aim of this study wa
s to investigate whether electrodes continue to function effectively beyond
100% consumption by measuring the pressure output of electrodes up to 300%
consumption or until electrode failure,
Materials and Methods: We assessed new, refurbished, and twice-refurbished
electrodes to compare their respective performances, Results: Rather than a
decrease, there was a trend toward increased pressure output beyond 100% e
lectrode consumption, with no drop in pressure prior to 300% consumption or
electrode failure. All three types of electrodes functioned adequately abo
ve 100% consumption.
Conclusion: These observations suggest that discarding the electrode before
the completion of the treatment or electrode failure is not warranted. Usi
ng a single electrode rather than multiple electrodes to complete a treatme
nt has the potential to save both time and cost.