Calibration of a polarized neutron reflectometer

Citation
H. Fredrikze et Rwe. Van De Kruijs, Calibration of a polarized neutron reflectometer, PHYSICA B, 297(1-4), 2001, pp. 143-147
Citations number
6
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science
Journal title
PHYSICA B
ISSN journal
09214526 → ACNP
Volume
297
Issue
1-4
Year of publication
2001
Pages
143 - 147
Database
ISI
SICI code
0921-4526(200103)297:1-4<143:COAPNR>2.0.ZU;2-8
Abstract
In a polarized neutron reflection experiment four reflectivities R-kl are d etermined, where k and l denote the spin state of the incoming and reflecte d neutrons, respectively, either parallel or anti-parallel to the guide fie lds along the incoming and reflected beam. A calibration of the setup is ne eded because the polarizer nor the analyzer are perfect and moreover the ef fective polarizing power of the polarizer and analyzer may depend on the no minal spin state of the incoming and reflected neutrons. It is shown that f our measurements without a sample and four measurements using a magneticall y saturated sample are needed to determine the five instrumental parameters that are relevant to interpret the results of a polarized neutron reflecti on experiment properly. For a time-of-flight reflectometer these parameters are wavelength dependent. (C) 2001 Elsevier Science B.V. All rights reserv ed.