A batch of 80 DL-alanine dosimeters was supplied to Hemocentro of the Hospi
tal and Clinics of Faculdade de Medicina de Ribeirao Preto (HC-FMRP) SP, Br
azil for the purpose of quality control of the radiation dose delivered to
blood bags. The irradiation was made using two (40 x 40) cm(2) parallel opp
osed radiation fields each with 80cm of source to surface distance in the R
adiotherapy Section of HC-FMRP with the Co-60 teletherapy unit. The calcula
ted radiation absorbed dose at the center of the box was 20 Gy. The dosimet
er readings were performed using a Varian E-4 ESR Spectrometer operating in
X-band. For the 80 dosimeters and over the irradiation volume throughout a
blood bag, the minimum and maximum doses were 14 and 23 Gy, respectively.
The mean dose was (18 +/- 2) Gy (1 sigma), and the coefficient of variabili
ty was 11.1%. Alanine dosimeters demonstrated easy handling, good precision
and adequate sensitivity for this application, (C) 2001 Elsevier Science L
td. All rights reserved.