A direct determination of extra virgin olive oil acidity by an automated fl
ow injection titration is proposed. The oil sample (280 mul) is directly in
jected, without dilution, in the carrier stream, a n-propanol solution flow
ing at 2 ml/min and containing potassium hydroxide 1 mM as titrant and phen
olphthalein 5 10(-5) M as indicator. In the mixing chamber(1.5 mi) the samp
le mixes and reacts with the carrier stream. By an optical fiber beam, conn
ected to a spectrophotometer set at 562 nm, we are able to appreciate conti
nuously the phenolphthalein absorbance (Abs) decrease. The area of the peak
"Delta Abs vs, time" has a linear correlation with the logarithm 1 of the
sample acidity. Several extra virgin olive oil samples have been analysed.
The correlation curve between the official (O) and proposed methods (P) is
P = (0.02 +/-0.02) + (1.02 +/-0.03)O, with r(2) = 0.997 and s(r) = 0.033; t
he accuracy and precision, expressed as mean error and RSD, are both <5%; t
he number of samples analysed per hour, the solvent and the sample consumpt
ion, in comparison with the official procedure, are respectively 12-60 vers
us 7-8, 2-10 ml versus 100-150 ml and 0.5 ml versus 10-20 ml. These feature
s and the simple procedure indicate how the proposed method could represent
an interesting alternative to the European community official method for t
he determination of free fatty acids in oil. <(c)> 2001 Elsevier Science Lt
d. All rights reserved.