J. Grobner et al., Intercomparison of aerosol optical depth measurements in the UVB using Brewer spectrophotometers and a Li-Cor spectrophotometer, GEOPHYS R L, 28(9), 2001, pp. 1691-1694
The first Iberian UV radiation intercomparison was held at "El Arenosillo"-
Huelva station of the Institute Nacional de Tecnica Aeroespatial (INTA) fr
om September 1 to 10, 1999. During this campaign, seven Brewer spectrophoto
meters and one Li-Cor spectrophotometers measured the total column aerosol
optical depth (AOD) at 306, 310, 313.5, 316.75 and 320 nm. The BOD calibrat
ion of one Brewers was transferred to all other Brewers using one day of in
tensive measurements. The remaining days were used to observe the stability
and reproducibility of the AOD measurements by the different instruments.
All Brewer spectrophotometers agreed to within an AOD of 0.03 during the wh
ole measurement campaign. The differences in AOD between the Li-Cor spectro
photometer and the Brewer spectrophotometers were between -0.07 and +0.02 a
t 313.5, 316.75, and 320 nm. This investigation demonstrates the possibilit
y of using the existing worldwide Brewer network as a global UV aerosol net
work for AOD monitoring.