Development of a 600 MHz wide bore(89mm) NMR system using internal tin wires

Citation
Wj. Shen et al., Development of a 600 MHz wide bore(89mm) NMR system using internal tin wires, IEEE APPL S, 11(1), 2001, pp. 2429-2432
Citations number
2
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science
Journal title
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY
ISSN journal
10518223 → ACNP
Volume
11
Issue
1
Year of publication
2001
Part
2
Pages
2429 - 2432
Database
ISI
SICI code
1051-8223(200103)11:1<2429:DOA6MW>2.0.ZU;2-5
Abstract
Employing high performance internal-tin Nb3Sn superconducting wire in NMR s uperconducting magnets, it is possible to reduce NMR magnet system cost. A 600 MHz NMR magnet system with 89mm room temperature bore has been designed , constructed and is being tested. The magnet and cyrostat were designed to meet existing 600 MHz NMR superconducting magnet standards. The design involved extensive mechanical analysis of stress/strain of the c oils and coil forms. Construction also involved the development of supercon ducting joints. The magnet, which is cooled bq: liquid helium at 4.2K, cons ists of three Nb3Sn coils and five NbTi coils. The operating current, store d energy, outer diameter and the height of the whole system are 87.5 A, 2 M J, 865 mm and 2020 mm, respectively. In this paper, an overview and analysi s of the design is presented and preliminary test results are reported.