An array of 7492 SINIS Josephson junctions was used to generate fast revers
ed de (FRDC) voltages of approximately 1.1 V for the study of thermoelectri
c effects in planar multifunction thermal converters (PMJTCs). A bias curre
nt source for fast reversal with a frequency of up to 10 kHz has been desig
ned. For the first time, results of Josephson measurements on two different
PMJTCs excellently agree with the results obtained using a semiconductor F
RDC source within 0.06 muV/V up to 200 Hz. At higher frequencies, a systema
tic slope in the relative FRDC-DC voltage transfer difference versus the re
versing frequency of about 10(-7)/kHz has been observed, which could not be
measured,vith the semiconductor source, This slope is probably caused by t
he rf-filters required in the cryoprobe.