A number of international comparisons of power meter calibrations at power
frequencies were conducted in the past, all of them being done at 120 V and
5 A. This comparison of power meter calibrations between the National Rese
arch Council of Canada and the National Institute of Standards and Technolo
gy, USA, was done to include voltage ranges up to 600 V, current ranges up
to 100 A, and at power factors of 1.0, 0.5 lead and lag, and 0.0 lead and l
ag. The comparison was implemented by a transfer standard consisting of a m
odified commercial time-division multiplier type wattmeter based on a devel
opment at the National Research Council of Canada. The results indicate tha
t there are no significant differences in the overall accuracy of ac power
measurements in each laboratory up to 600 V, 100 A, at all power factors.