NRC-NIST comparison of power meter calibrations at 60 Hz and ranges up to 600 V, 100 A

Citation
E. So et al., NRC-NIST comparison of power meter calibrations at 60 Hz and ranges up to 600 V, 100 A, IEEE INSTR, 50(2), 2001, pp. 353-355
Citations number
7
Categorie Soggetti
Instrumentation & Measurement
Journal title
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
ISSN journal
00189456 → ACNP
Volume
50
Issue
2
Year of publication
2001
Pages
353 - 355
Database
ISI
SICI code
0018-9456(200104)50:2<353:NCOPMC>2.0.ZU;2-D
Abstract
A number of international comparisons of power meter calibrations at power frequencies were conducted in the past, all of them being done at 120 V and 5 A. This comparison of power meter calibrations between the National Rese arch Council of Canada and the National Institute of Standards and Technolo gy, USA, was done to include voltage ranges up to 600 V, current ranges up to 100 A, and at power factors of 1.0, 0.5 lead and lag, and 0.0 lead and l ag. The comparison was implemented by a transfer standard consisting of a m odified commercial time-division multiplier type wattmeter based on a devel opment at the National Research Council of Canada. The results indicate tha t there are no significant differences in the overall accuracy of ac power measurements in each laboratory up to 600 V, 100 A, at all power factors.