An international key comparison of effective efficiency was carried out fro
m 1996 to 1998 between four National Metrology Institutes (NMIs) at 62 GHz
in IEC R620 rectangular waveguide standard, The traveling item was a wavegu
ide temperature compensated thermistor mount. The following NMIs participat
ed: National Physical Laboratory (NPL), Malvern, U,K,; State Scientific Ind
ustrial Association (SSIA), Kharkov, Ukraine; National Institute of Standar
ds and Teehnology (NIST), Boulder, CO; and Bureau National de Metrologie-La
boratoire Central des Industries Electriques (BNM-LCIE), Fontenay-aux-Roses
, France, which acted as the pilot laboratory for the GT-RF/95-1 key compar
ison. Good agreement was found among the results from the different NMIs, w
ith all results agreeing within one standard uncertainty (kappa = 1), which
ranged from approximately 0.5% to 0.9%.