An original, rapid and sensitive high-performance liquid chromatographic (H
PLC) method has been developed for the detection of 5, 10, 15, 20 tetra-mes
o-hydroxyphenylchlorine (m-THPC), a photosensitizer used in the photodynami
c therapy (PDT) of cancers. Chromatographic separation was carried but on a
C-8 Zorbax(R) column (80 x 4 mm, 5 mum). The mobile phase was an ethanol/a
queous sulphuric acid, pH 2.0 (65/35 v/v), in an isocratic mode yielding to
rapid analysis (3.1 min) with narrow peaks. As the fluorescence intensity
was found to be highly pH-dependent and to increase with pH values, a post-
column device prior to the fluorescence detection (lambda (exc) = 423 nm, l
ambda (em) = 650 nm) was used to allow the addition of a 0.05 mol/L Na2HPO4
solution to the mobile phase. Compared to standard conditions, a 300% incr
ease of the fluorescence intensity was obtained for optimized operating con
ditions using experimental design. The validation of this analytical method
showed;hat the response function was linear for concentrations up to 1000
mug/L (1.47 x 10(-6) mol/L) with a detection limit of 188 pg (S/N = 3). Cop
yright (C) 2001 John Wiley & Sons, Ltd.