Calibration of the Preston probe for high Reynolds number flows

Citation
Mv. Zagarola et al., Calibration of the Preston probe for high Reynolds number flows, MEAS SCI T, 12(4), 2001, pp. 495-501
Citations number
24
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
MEASUREMENT SCIENCE & TECHNOLOGY
ISSN journal
09570233 → ACNP
Volume
12
Issue
4
Year of publication
2001
Pages
495 - 501
Database
ISI
SICI code
0957-0233(200104)12:4<495:COTPPF>2.0.ZU;2-B
Abstract
Experiments were performed in order to calibrate Preston probes for high Re ynolds number applications. The results extend previous calibrations to Rey nolds numbers that are approximately 30 times larger. The new calibration i s valid in the range 6.4 < x* < 11.3, that is, 4.3 < y* <less than> 8.7 and 280 < D+ < 45 000. Within this range, the wall shear stress may be found t o within better than +/-0.8%, as long as inner layer scaling applies up to a wall distance equal to the probe diameter.