E. Moreno-barriuso et al., Comparing laser ray tracing, the spatially resolved refractometer, and theHartmann-Shack sensor to measure the ocular wave aberration, OPT VIS SCI, 78(3), 2001, pp. 152-156
Purpose. To compare quantitatively three techniques to measure the optical
aberrations of the human eye: laser ray tracing (LRT), the Hartmann-Shack w
avefront sensor (H-S), and the spatially resolved refractometer (SRR), LRT
and H-S are objective imaging techniques, whereas SRR is psychophysical, Me
thods. Wave aberrations were measured in two normal subjects with all three
techniques implemented in two different laboratories. Results. We compared
the experimental variability of the results obtained with each technique w
ith the overall Variability across the three methods. For the two subjects
measured (RMS wavefront error 0.5 mum and 0.9 mum, respectively), we found
a close agreement; the average standard deviation of the Zernike coefficien
ts within a given method was 0.07 mum, whereas the average global standard
deviation across techniques was 0.09 mum, which is only slightly higher, Co
nclusions. There is a close match between the Zernike coefficients obtained
by LRT, H-S, and SRR. Thus, all three techniques provide similar informati
on concerning wave aberration when applied to normal human eyes. However, t
he methods are operationally different, and each has advantages and disadva
ntages depending on the particular application.