On-line preconcentration system for bismuth determination in urine by flowinjection hydride generation inductively coupled plasma atomic emission spectrometry

Citation
S. Moyano et al., On-line preconcentration system for bismuth determination in urine by flowinjection hydride generation inductively coupled plasma atomic emission spectrometry, TALANTA, 54(2), 2001, pp. 211-219
Citations number
32
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
TALANTA
ISSN journal
00399140 → ACNP
Volume
54
Issue
2
Year of publication
2001
Pages
211 - 219
Database
ISI
SICI code
0039-9140(20010412)54:2<211:OPSFBD>2.0.ZU;2-I
Abstract
An on-line bismuth preconcentration and determination system implemented wi th hydride generation inductively coupled plasma atomic emission spectromet ry (HG-ICP-AES) associated to flow injection (FI) was studied. Quinolin-8-o l and Amberlite XAD-7 were used for the retention of bismuth, at FH 4.5. Th e bismuth complex was removed from the micro-column with nitric acid. The d etection limit value for the preconcentration of 100 mi of aqueous solution was 0.02 ng ml(-1) with a relative standard deviation (R.S.D.) of 3.5%, ca lculated from the peak heights obtained. The calibration graph using the pr econcentration system for bismuth was linear with a correlation coefficient of 0.999 at levels near the detection limits up to at Least 100 ng ml(-1). The method was successfully applied to the determination of bismuth in hum an urine samples. (C) 2001 Elsevier Science B.V, All rights reserved.