Jl. Ren et al., Improved fluorimetric determination of dissolved aluminium by micelle-enhanced lumogallion complex in natural waters, ANALYST, 126(5), 2001, pp. 698-702
A modification of the aluminium-lumogallion fluorescence measurement in the
presence of the non-ionic surfactant Triton X-100 is presented. The detect
ion limit for dissolved Al is 0.7 nM, with a relative standard deviation of
3.6% at an Al level of 5.0 nM. Compared with previously reported methods i
n the literature, the method described here is free from matrix effects and
can be used for the determination of aluminium in fresh, estuarine and sal
ine waters. The interferences from iron and fluoride were minimized by the
addition of o-phenanthroline and Be2+, respectively. The analysis of NIST S
RM 1643C and PRC standard 2430101 by the proposed method provides results c
onsistent with the certified values. A successful inter-laboratory calibrat
ion exercise also demonstrates the merit of the proposed method for the det
ermination of Al in environmental and marine sciences.