Development of a new flow-through bulk optode for the determination of manganese(II)

Citation
C. Sanchez-pedreno et al., Development of a new flow-through bulk optode for the determination of manganese(II), FRESEN J AN, 369(7-8), 2001, pp. 680-683
Citations number
20
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY
ISSN journal
09370633 → ACNP
Volume
369
Issue
7-8
Year of publication
2001
Pages
680 - 683
Database
ISI
SICI code
0937-0633(200104)369:7-8<680:DOANFB>2.0.ZU;2-D
Abstract
A flow-through bulk optode based on the use of 1-(2-pyridylazo)-2-naphthol (PAN) immobilized in a plasticized poly(vinyl chloride) membrane entrapped in a cellulose support, in conjuntion with the flow injection analysis tech nique, is proposed for the determination of manganese(II). The calibration graph obtained at 570 nm was linear in the range 0.27-27.5 mg L-1 (5 x 10(- 6)-5;( 10(-4) M) MN(II) with a detection limit of 0.18 mg L-1. The coeffici ents of variation of the sensor response for 5.5 mg L-1 of Mn(II) were +/-0 .22% for consecutive measurements (n = 10). +/-0.48% between days in = 5) a nd +/-0.38% between different membranes (n = 6). The sensor was readily reg enerated with the carrier acetic acid/acetate buffer of pH 4.5. The method was applied to the determination of manganese in steels, waters and lemon t ree leaves.