Cnh. Enzweiler et al., Rib artifacts in electron beam tomography: Incidence and severity without and with the cone beam reconstruction algorithm, J COMPUT AS, 25(3), 2001, pp. 365-370
Citations number
27
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Electron beam tomography (EBT) may he compromised by rib artifacts. Two hun
dred forty-seven abdominal studies were performed without (Group A, n = 222
) or with (Group B. n = 25) the cone beam algorithm. One hundred eighty-six
(83.8%) and nine (36%) studies of Groups A and B, respectively, displayed
some level of artifact. In Groups A and B, major, minor, and no artifacts w
ere found in 115 (51.5%) and 0 (0%), 71 (32.3%) and 9 (36%), and 36 (16.2%)
and 16 (64%) patients, respectively (p < 0.01). The cone beam algorithm im
proves EBT studies of the abdomen.