Wear simulation of total hip prostheses with polyethylene against CoCr, alumina and diamond-like carbon

Citation
V. Saikko et al., Wear simulation of total hip prostheses with polyethylene against CoCr, alumina and diamond-like carbon, BIOMATERIAL, 22(12), 2001, pp. 1507-1514
Citations number
26
Categorie Soggetti
Multidisciplinary
Journal title
BIOMATERIALS
ISSN journal
01429612 → ACNP
Volume
22
Issue
12
Year of publication
2001
Pages
1507 - 1514
Database
ISI
SICI code
0142-9612(200106)22:12<1507:WSOTHP>2.0.ZU;2-A
Abstract
The wear of ultra-high molecular weight polyethylene acetabular cups was st udied with a new biaxial hip wear simulator using diluted calf serum as a l ubricant. The cups had been packed and gamma-irradiated in argon. The cups were articulated against two established types of femoral head, alumina and CoCr, and one experimental type, CoCr coated with diamond-like carbon (DLC ). The diameter of the heads was 28 mm. Polyethylene against alumina and ag ainst CoCr were studied because their clinical wear behaviour is relatively well known. The new simulator was validated with these established materia ls. The wear mechanisms, including the size and shape of the wear particles , agreed well with those seen in clinical studies. The average wear rates o f the cups against alumina and CoCr heads were 48 and 56 mg per 1 million c ycles, respectively. The order is in agreement with clinical observations. The average wear rate against DLC was 58 mg per 1 million cycles. As a coun terface for polyethylene, DLC did not markedly differ from alumina and CoCr , (C) 2001 Elsevier Science Ltd. All rights reserved.