FROST: a low-noise high-rate photon counting ASIC for X-ray applications

Citation
M. Prest et al., FROST: a low-noise high-rate photon counting ASIC for X-ray applications, NUCL INST A, 461(1-3), 2001, pp. 435-439
Citations number
4
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
461
Issue
1-3
Year of publication
2001
Pages
435 - 439
Database
ISI
SICI code
0168-9002(20010401)461:1-3<435:FALHPC>2.0.ZU;2-F
Abstract
FRONTier RADiography is an R&D project to assess the feasibility of digital mammography with Synchrotron Radiation at the ELETTRA Light Source in Trie ste. In order to reach an acceptable time duration of the exam, a fast-and low-noise photon counting ASIC has been developed in collaboration with Aur elia Microelettronica, called Frontrad ReadOut SysTem. It is a multichannel counting system, each channel being made of a low-noise charge-sensitive p reamplifier optimized for X-lay energy range (10 100 keV), a CR-RC2 shaper. a discriminator and a 16-bit counter. In older to set the discriminator th reshold, a set of a global 6-bit DAC and a local (per channel) 3-bit DAC ha s been implemented within the ASIC. We report on the measurements done with the 8-channel prototype chip and the comparison with the simulation result s. (C) 2001 Elsevier Science B.V. All rights reserved.