Paradoxical induction of cross-talk in a dual chamber pacemaker by increasing the ventricular output

Authors
Citation
Ss. Barold, Paradoxical induction of cross-talk in a dual chamber pacemaker by increasing the ventricular output, PACE, 24(5), 2001, pp. 906-907
Citations number
2
Categorie Soggetti
Cardiovascular & Respiratory Systems","Cardiovascular & Hematology Research
Journal title
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
ISSN journal
01478389 → ACNP
Volume
24
Issue
5
Year of publication
2001
Pages
906 - 907
Database
ISI
SICI code
0147-8389(200105)24:5<906:PIOCIA>2.0.ZU;2-#
Abstract
Cross-talk was induced in a unipolar DDD pacemaker by increasing the output of the ventricular channel. Oversensing by the ventricular channel was pro bably due to the detection of a relatively large summation signal consistin g of the afterpotential and the superimposed activity from the atrial stimu lus.