H. Alheit et al., Patient position reproducibility in fractionated stereotactically guided conformal radiotherapy using the BrainLab (R) mask system, STRAH ONKOL, 177(5), 2001, pp. 264-268
Purpose: Dedicated mask systems nowadays allow the use of stereotactic radi
otherapy in fractionated regimes, therefore combining the advantages of hig
h precision radiotherapy with the biological benefit of fractionation. Ther
efore the knowledge of institution specific isocenter accuracy is essential
for decision-making about margins to be allowed to form the planning targe
t volume.
Patients and Method: Measurements of isocenter deviations during fractionat
ed treatments were performed in 33 patients using the simulator Simulix-xy
(Oldelft) in connection with the BrainLab(R) angiographic Localizer-box as
well as port-films. In both cases repeated images were overlaid by use of a
natomical Landmarks with a methodical accuracy in the order of 0.5 mm.
Results: Both methods yield random isocenter deviations of Less then 2 mm (
standard deviation) in all three directions and no significant systematic d
eviations. These values are in the order of the accuracy of the method, obt
ained by comparison of two independent investigators, as well as they are c
omparable with the literature.
Conclusions: The accuracy of Less than 2 mm indicates safety margins of 3-4
mm as sufficient for clinical routine to cover the target in 95.5% of all
set-ups (2 SD).