S. Miljanic et M. Ranogajeckomor, APPLICATION OF CAVITY THEORY TO THE RESPONSE OF VARIOUS TLDS TO CO-60GAMMAS DEGRADED IN WATER, Physics in medicine and biology, 42(7), 1997, pp. 1335-1349
Several types of thermoluminescent dosimeters (LiF:Mg,Ti, Al2O3:Mg,Y a
nd CaF2:Mn) were irradiated at different depths in a water phantom pla
ced at a distance of 2.5 m from a panoramic Co-60 source. Detectors we
re encapsulated in Plexiglas holders with a wall thickness of 0.5 cm.
Reference dosimetry was carried out using a Fricke dosimeter and an io
nization chamber. The experimental data were compared with the predict
ions of the general cavity theory for gamma ray spectra at different d
epths of water. The suitability of parameters of the cavity theory pro
posed by different authors was evaluated in the analysis of the experi
mental results. The results show that there is no need for any modific
ation to the original and simple Burlin expression which gives very go
od agreement with the experimental values.