Mg. Herman et al., Clinical use of electronic portal imaging: Report of AAPM radiation therapy committee Task Group 58, MED PHYS, 28(5), 2001, pp. 712-737
Citations number
140
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
AAPM Task Group 58 was created to provide materials to help the medical phy
sicist and colleagues succeed in the clinical implementation of electronic
portal imaging devices (EPIDs) in radiation oncology. This complex technolo
gy has matured over the past decade and is capable of being integrated into
routine practice. However, the difficulties encountered during the specifi
cation, installation, and implementation process can be overwhelming. TG58
was charged with providing sufficient information to allow the users to ove
rcome these difficulties and put EPIDs into routine clinical practice. In a
nswering the charge, this report provides; comprehensive information about
the physics and technology of currently available EPID systems; a detailed
discussion of the steps required for successful clinical implementation, ba
sed on accumulated experience; a review of software tools available and cli
nical use protocols to enhance EPID utilization; and specific quality assur
ance requirements for initial and continuing clinical use of the systems. S
pecific recommendations are summarized to assist the reader with successful
implementation and continuing use of an EPID, (C) 2001 American Associatio
n of Physicists in Medicine.