Algorithms for fast axisymmetric drop shape analysis measurements by a charge coupled device video camera and simulation procedure for test and evaluation

Citation
L. Busoni et al., Algorithms for fast axisymmetric drop shape analysis measurements by a charge coupled device video camera and simulation procedure for test and evaluation, REV SCI INS, 72(6), 2001, pp. 2784-2791
Citations number
19
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
REVIEW OF SCIENTIFIC INSTRUMENTS
ISSN journal
00346748 → ACNP
Volume
72
Issue
6
Year of publication
2001
Pages
2784 - 2791
Database
ISI
SICI code
0034-6748(200106)72:6<2784:AFFADS>2.0.ZU;2-9
Abstract
A set of fast algorithms for axisymmetric drop shape analysis measurements is described. Speed has been improved by more than 1 order of magnitude ove r previously available procedures. Frame analysis is performed and drop cha racteristics and interfacial tension gamma are computed in less than 40 ms on a Pentium III 450 MHz PC, while preserving an overall accuracy in Delta gamma/gamma close to 1x10(-4). A new procedure is described to evaluate bot h the algorithms performance and the contribution of each source of experim ental error to the overall measurement accuracy. (C) 2001 American Institut e of Physics.