The practical application of colour-coded duplex sonography shows that
physico-technical artifacts of colour Doppler imaging may not be obvi
ous in all cases and can lead to misinterpretation. Slice-thickness ar
tifacts are due to restricted transversal resolution, as in the B-mode
, and may cause false negative results in vascular occlusions. Specifi
c problems in creating and processing the colour flow image interfere
with lateral resolution and lead to errors in the quantification of st
enoses from the colour Doppler image. Mirror image artifacts produce p
hantom pictures that can be easily identified as such in some cases, b
ut may resemble recesses in the arterial wall in other cases. Phenomen
a of sound shadowing resulting from the same reasons as in B-mode may
lead to assume vascular occlusions in cases of long-range stenoses. Ar
tifacts of the sonication angle in case of blood vessel tortuosity may
lead to misdiagnosis of a non-existent retrograde flow. The artifacts
mentioned do restrict the capacity of colour coded duplex sonography
for localising flow in the ultrasound image. Users of the method shoul
d therefore be familiar with these artifacts.